Comparative study of i-SENS glucometers in neonates using capillary blood samples

  • Ha Nui Kim and Soo-Young Yoon EMAIL logo

Abstract

Objectives

The accuracy of point-of-care blood glucometers in the detection and evaluation of neonatal hypoglycemia is a concern. This study compared the performance of three i-SENS glucometers with that of the YSI 2300 STAT Plus Analyzer, which was used as a reference.

Methods

The leftover neonatal capillary blood samples of 319 patients were used in this study. The evaluation process and accuracy performance criteria were based on the International Organization for Standardization 15197:2013 guidelines. The evaluation involved three i-SENS glucometers (BAROzen H Expert plus, CareSens PRO, and CareSens H Beat) and the ACCU-CHEK® Inform II glucometer.

Results

The accuracy evaluation yielded acceptable results as follows: a) 100 and 100% for BAROzen H Expert plus; 99.8 and 100% for CareSens PRO; 98.7%, and 97.2% for CareSens H Beat glucometers were within the range of ±0.8 mmol/L (15 mg/dL) and ±15% of the average reference values at glucose concentrations <5.55 mmol/L (100 mg/dL) and ≥5.55 mmol/L (100 mg/dL), respectively, and b) all estimated glucose values (100%) were within the zones A and B of Consensus Error Grid for all three i-SENS glucometers. There was good correlation between the glucose values estimated by the glucometers and the reference values (R>0.990).

Conclusions

This study demonstrated that i-SENS glucometers exhibit acceptable performance and can be used as effective point-of-care devices in neonates.


Corresponding author: Soo-Young Yoon, MD, PhD, Department of Laboratory Medicine, Korea University Guro Hospital, 148, Gurodong-ro, Guro-gu, Seoul 08308, Republic of Korea, Phone: +82 2 2626 3246, Fax: +82 2 2626 1465, E-mail:

Funding source: Korean Health Technology R&D Project

Funding source: Ministry of Health & Welfare, Republic of Korea

Award Identifier / Grant number: H15C1512

Funding source: i-SENS, Inc.

Funding source: Korea Health Technology R&D Project through the Korea Health Industry Development Institute (KHIDI), funded by the Ministry of Health & Welfare, Republic of Korea

Award Identifier / Grant number: HR20C0021

Acknowledgments

We would like to thank Editage (www.editage.co.kr) for the English language editing and also i-SENS, Inc., for providing reagents with technical support.

  1. Research funding: This study was supported by a grant of the Korea Health Technology R&D Project through the Korea Health Industry Development Institute (KHIDI), funded by the Ministry of Health & Welfare, Republic of Korea (grant number: HR20C0021) and the Korean Health Technology R&D Project, Ministry of Health & Welfare, Republic of Korea (H15C1512) and i-SENS, Inc.

  2. Author contributions: HK and SY reviewed the literature and conceived the study. SY was involved in protocol development, obtaining ethical approval, patient recruitment and data analysis. HK wrote the first draft of the manuscript. All authors reviewed and edited the manuscript and approved the final version of the manuscript.

  3. Competing interests: Authors state no conflict of interest.

  4. Informed consent: Informed consent was obtained from all individuals included in this study.

  5. Ethical approval: The research related to human use has been complied with all the relevant national regulations, institutional policies and in accordance the tenets of the Helsinki Declaration, and has been approved by Institutional Review Board of Korea University Guro Hospital (IRB No: 2019GR0016).

References

1. Roth-Kleiner, M, Stadelmann Diaw, C, Urfer, J, Ruffieux, C, Werner, D. Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting. Eur J Pediatr 2010;169:1387–95.10.1007/s00431-010-1243-2Search in Google Scholar PubMed

2. Lucas, A, Morley, R, Cole, TJ. Adverse neurodevelopmental outcome of moderate neonatal hypoglycaemia. BMJ 1988;297:1304–8.10.1136/bmj.297.6659.1304Search in Google Scholar PubMed PubMed Central

3. Nuntnarumit, P, Chittamma, A, Pongmee, P, Tangnoo, A, Goonthon, S. Clinical performance of the new glucometer in the nursery and neonatal intensive care unit. Pediatr Int 2011;53:218–23.10.1111/j.1442-200X.2010.03214.xSearch in Google Scholar PubMed

4. Movalia, MK, Ogino, MT. Point of care glucose testing in neonatal hypoglycemia. Point Care 2006;5:95–9.10.1097/01.poc.0000232027.69785.f5Search in Google Scholar

5. Ho, H, Yeung, W, Young, B. Evaluation of “point of care” devices in the measurement of low blood glucose in neonatal practice. Arch Dis Child Fetal Neonatal Ed 2004;89:F356–9.10.1136/adc.2003.033548Search in Google Scholar PubMed PubMed Central

6. Freckmann, G, Jendrike, N, Baumstark, A, Pleus, S, Liebing, C, Haug, C. User performance evaluation of four blood glucose monitoring systems applying ISO 15197:2013 accuracy criteria and calculation of insulin dosing errors. Diabetes Ther 2018;9:683–97.10.1007/s13300-018-0392-6Search in Google Scholar PubMed PubMed Central

7. Solnica, B, Naskalski, J, Gernand, W, Sieradzki, J. Evaluation of glucose meters performance in detection hypoglycemia. Point Care 2008;7:144.10.1097/01.POC.0000335850.55143.5dSearch in Google Scholar

8. Ekhlaspour, L, Mondesir, D, Lautsch, N, Balliro, C, Hillard, M, Magyar, K, et al.. Comparative accuracy of 17 point-of-care glucose meters. J Diabetes Sci Technol 2017;11:558–66. https://doi.org/10.1177/1932296816672237.Search in Google Scholar PubMed PubMed Central

9. Klonoff, DC. The need for separate performance goals for glucose sensors in the hypoglycemic, normoglycemic, and hyperglycemic ranges. Diabetes Care 2004;27:834–6.10.2337/diacare.27.3.834Search in Google Scholar PubMed

10. Freckmann, G, Schmid, C, Baumstark, A, Rutschmann, M, Haug, C, Heinemann, L. Analytical performance requirements for systems for self-monitoring of blood glucose with focus on system accuracy: relevant differences among ISO 15197:2003, ISO 15197:2013, and current FDA recommendations. J Diabetes Sci Technol 2015;9:885–94.10.1177/1932296815580160Search in Google Scholar PubMed PubMed Central

11. Rodin, D, Kirby, M, Sedogin, N, Shapiro, Y, Pinhasov, A, Kreinin, A. Comparative accuracy of optical sensor-based wearable system for non-invasive measurement of blood glucose concentration. Clin Biochem 2019;65:15–20.10.1016/j.clinbiochem.2018.12.014Search in Google Scholar PubMed

12. Ginsberg, BH. Factors affecting blood glucose monitoring: sources of errors in measurement. J Diabetes Sci Technol 2009;3:903–13.10.1177/193229680900300438Search in Google Scholar PubMed PubMed Central

13. Stene, D, Song, L. Investigation of maltose interference on the Roche ACCU-CHEK Inform II blood glucose meter. Point Care 2016;15:26–9.10.1097/POC.0000000000000083Search in Google Scholar

14. Chakraborty, PP, Patra, S, Bhattacharjee, R, Chowdhury, S. Erroneously elevated glucose values due to maltose interference in mutant glucose dehydrogenase pyrroloquinolinequinone (mutant GDH-PQQ) based glucometer. BMJ Case Rep 2017;2017:bcr-2017-219928. https://doi.org/10.1136/bcr-2017-219928.Search in Google Scholar PubMed PubMed Central

15. Klonoff, DC, Parkes, JL, Kovatchev, BP, Kerr, D, Bevier, WC, Brazg, RL, et al.. Investigation of the accuracy of 18 marketed blood glucose monitors. Diabetes Care 2018;41:1681–8.10.2337/dc17-1960Search in Google Scholar PubMed

16. Thompson-Branch, A, Havranek, T. Neonatal hypoglycemia. Pediatr Rev 2017;38:147–57.10.1542/pir.2016-0063Search in Google Scholar PubMed

17. Guemes, M, Rahman, SA, Hussain, K. What is a normal blood glucose? Arch Dis Child 2016;101:569–74.10.1136/archdischild-2015-308336Search in Google Scholar PubMed

18. Rafai, N, Horvath, AR, Wittwer, CT. Tietz textbook of clinical chemistry and molecular diagnostics. St. Louis, Missouri: Elsevier Health Sciences; 2018.Search in Google Scholar

19. Morales, J, Schneider, D. Hypoglycemia. Am J Med 2014;127:S17–24.10.1016/j.amjmed.2014.07.004Search in Google Scholar PubMed

20. Alsaleem, M, Saadeh, L, Kamat, D. Neonatal hypoglycemia: a review. Clin Pediatr 2019;58:1381–6.10.1177/0009922819875540Search in Google Scholar PubMed

21. Stanley, CA, Rozance, PJ, Thornton, PS, De Leon, DD, Harris, D, Haymond, MW, et al.. Re-evaluating “transitional neonatal hypoglycemia”: mechanism and implications for management. J Pediatr-Us 2015;166:1520–5. https://doi.org/10.1016/j.jpeds.2015.02.045.Search in Google Scholar PubMed PubMed Central

22. Kumar, R, Madan, A, Benitz, WE, Widness, JA. Reduction in red blood cell transfusions using a bedside analyzer in extremely low birth weight infants. Pediatr Res 2004;55:21–5. https://doi.org/10.1038/sj.jp.7211201.Search in Google Scholar PubMed

23. Sacks, DB, Arnold, M, Bakris, GL, Bruns, DE, Horvath, AR, Kirkman, MS, et al.. Executive summary: guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem 2011;57:793–8.10.1373/clinchem.2011.163634Search in Google Scholar PubMed

24. Cho, J, Ahn, S, Yim, J, Cheon, Y, Jeong, SH, Lee, SG, et al.. Influence of vitamin C and maltose on the accuracy of three models of glucose meters. Ann Lab Med 2016;36:271–4.10.3343/alm.2016.36.3.271Search in Google Scholar PubMed PubMed Central

25. Dogan, K, Kayalp, D, Ceylan, G, Azak, A, Senes, M, Duranay, M, et al.. Falsely elevated glucose concentrations in peritoneal dialysis patients using icodextrin. J Clin Lab Anal 2016;30:506–9.10.1002/jcla.21887Search in Google Scholar PubMed PubMed Central

26. Ramel, SE, Long, JD, Gray, H, Durrwachter-Erno, K, Demerath, EW, Rao, R. Neonatal hyperglycemia and diminished long-term growth in very low birth weight preterm infants. J Perinatol 2013;33:882–6.10.1038/jp.2013.77Search in Google Scholar PubMed

27. Simsek, DG, Ecevit, A, Hatipoglu, N, Coban, A, Arisoy, AE, Bas, F, et al.. Neonatal hyperglycemia, which threshold value, diagnostic approach and treatment? Turkish Neonatal and Pediatric Endocrinology and Diabetes Societies consensus report. Turk Pediatr Arsivi 2018;53:S234–8.10.5152/TurkPediatriArs.2018.01821Search in Google Scholar PubMed PubMed Central

Received: 2020-09-09
Accepted: 2021-01-07
Published Online: 2021-01-22
Published in Print: 2021-05-26

© 2021 Walter de Gruyter GmbH, Berlin/Boston

Downloaded on 25.4.2024 from https://www.degruyter.com/document/doi/10.1515/cclm-2020-1367/html
Scroll to top button