Evaluation of four commercial, fully automated SARS-CoV-2 antibody tests suggests a revision of the Siemens SARS-CoV-2 IgG assay

  • Christian Irsara ORCID logo , Alexander E. Egger , Wolfgang Prokop , Manfred Nairz , Lorin Loacker , Sabina Sahanic , Alex Pizzini , Thomas Sonnweber , Wolfgang Mayer , Harald Schennach , Judith Loeffler-Ragg , Rosa Bellmann-Weiler , Ivan Tancevski , Günter Weiss , Markus Anliker , Andrea Griesmacher and Gregor Hoermann EMAIL logo

Abstract

Objectives

Serological tests detect antibodies against Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) in the ongoing coronavirus disease-19 (COVID-19) pandemic. Independent external clinical validation of performance characteristics is of paramount importance.

Methods

Four fully automated assays, Roche Elecsys Anti-SARS-CoV-2, Abbott SARS-CoV-2 IgG, Siemens SARS-CoV-2 total (COV2T) and SARS-CoV-2 IgG (COV2G) were evaluated using 350 pre-pandemic samples and 700 samples from 245 COVID-19 patients (158 hospitalized, 87 outpatients).

Results

All tests showed very high diagnostic specificity. Sensitivities in samples collected at least 14 days after disease onset were slightly lower than manufacturers’ claims for Roche (93.0%), Abbott (90.8%), and Siemens COV2T (90.3%), and distinctly lower for Siemens COV2G (78.8%). Concordantly negative results were enriched for immunocompromised patients. ROC curve analyses suggest a lowering of the cut-off index for the Siemens COV2G assay. Finally, the combination of two anti-SARS-CoV-2 antibody assays is feasible when considering borderline reactive results.

Conclusions

Thorough on-site evaluation of commercially available serologic tests for detection of antibodies against SARS-CoV-2 remains imperative for laboratories. The potentially impaired sensitivity of the Siemens COV2G necessitates a switch to the company’s newly filed SARS-CoV-2 IgG assay for follow-up studies. A combination of tests could be considered in clinical practice.

Introduction

Coronavirus disease 2019 (COVID-19) is caused by the Severe Acute Respiratory Syndrome Virus 2 (SARS-CoV-2) [1], [2], and was first described in China in December 2019 and declared pandemic by the WHO on March 11, 2020 [3]. As of December 23, 2020 more than 78 million confirmed cases and more than 1.7 million deaths have been reported worldwide (for Austria: more than 342,000 confirmed cases and more than 5,500 deaths) [4]. The total prevalence is estimated to be higher due to unrecognized infections [5]. The gold standard for the primary diagnosis of acute SARS-CoV-2 infection remains the specific detection of viral RNA by reverse transcription polymerase chain reaction (RT-PCR) [6], [7]. However, molecular methods have some limitations, in particular the relatively short time frame of detectability post symptom onset and thus possible false negative results in patients who present with low viral load at later stages of the disease course [8], [9], [10]. In contrast, the presence of SARS-CoV-2 antibodies indicates a recent or prior exposure to the virus [11]. Serological tests are therefore important to assess the seroprevalence for monitoring the epidemiology of SARS-CoV-2 infection at the population-level [12]. Furthermore, serologic testing may aid to increase the accuracy of diagnosis in addition to PCR [8], [13], to select eligible donors for convalescent plasma [14], and to estimate the immunological response to vaccination [15]. The durability of the antibody response as well as the extent and duration of immunity against reinfection with SARS-CoV-2 are still under investigation [16].

Immune response to SARS-CoV-2 includes cell-mediated [17] and antibody-mediated immunity [13]. The spike (S) protein with its receptor-binding domain (RBD) and the nucleocapsid (N) protein are widely used as antigens to assess the humoral immune responses by detection of specific antibodies [18], [19], [20]. Serological assays currently used include neutralizing antibody assays, enzyme-linked immunosorbent assays (ELISA), automated chemiluminescence immunoassays (CLIA) and lateral flow immunoassays (LFIA) [19], [21]. Distinct serologic assays detect different antibody isotypes: IgG, IgM, IgA or all isotypes simultaneously (“total antibodies”).

A rapidly growing number of commercial SARS-CoV-2 antibody assays is available. Whereas the manufacturer of an in vitro diagnostic (IVD) device is obliged to declare performance specifications of the test, the need for independent validation of commercial assays in clinical settings has been highlighted in systematic reviews of the literature [22], [23]. On the one hand, medical laboratories have to verify that they meet the performance specifications. On the other hand, small sample sizes and a lack of samples from patients with mild to moderate clinical course represent a potential bias in performance studies [24], [25]. Our study aimed to evaluate four SARS-CoV-2 antibody assays on three fully automated large-scale laboratory analyzers manufactured by Abbott, Roche, and Siemens, respectively. To or knowledge this is the first published external validation of the Siemens SARS-CoV-2 IgG (COV2G) antibody test.

Materials and methods

Patients and study design

The present study was performed at the University Hospital of Innsbruck as part of the clinical evaluation of different SARS-CoV-2 serologic assays. All procedures performed in the present study involving human participants were in accordance with the ethical standards of the Institutional and/or National Research Committee and with the 1964 Helsinki declaration and its later amendments and were approved by the Ethics Committee of the Medical University of Innsbruck (Ethics Commission numbers: 1103/2020, 1167/2020).

A total of 245 patients with RT-PCR confirmed SARS-CoV-2 infection were included: hospitalized COVID-19 patients at the University Hospital of Innsbruck, reconvalescent COVID-19 patients with persistent cardio-pulmonary damage participating in a prospective observational study (CovILD-study, ClinicalTrials.gov number, NCT04416100) and reconvalescent persons volunteering as plasma donors at the Central Institute for Blood Transfusion and Immunology. According to the clinical course, patients were grouped as outpatients, hospitalized patients at the general ward, or hospitalized patients at the intensive care unit (ICU) ward, largely overlapping with the superordinated categories of mild, moderate and severe disease of the WHO scale [26]. In total, 700 patient samples were assessed. 75 patients had one, 66 patients two, 34 patients three, 23 patients four, 24 patients five and 23 patients six or more blood draws. Disease onset was defined as onset of clinical symptoms compatible with COVID-19 infection as determined by a questionnaire in convalescent donors and by reviewing individual health records in the other patients. If the patient was asymptomatic or the date of symptom onset was not available (n=35, 15.2%; corresponding to 81 samples, 11.6%), the date of the first positive SARS-CoV-2 RT-PCR was used instead. The median time span between symptom onset and RT-PCR-based diagnosis was five days. The patients’ characteristics are shown in Table 1.

Table 1:

Characteristics of COVID-19 patients.

Total Outpatient Hospitalized

General ward
Hospitalized

Intensive care
Absolute % Absolute % Absolute % Absolute %
Number
 n 245 100% 87 36% 116 47% 42 17%
Age, years
 Min 20 20 25 44
 Max 95 83 95 82
 Median (IQR) 56 (44–69) 39 (29–56) 64 (54–77) 59 (54–67)
Sex
 Female 87 36% 34 39% 42 36% 11 26%
 Male 158 64% 53 61% 74 64% 31 74%
Symptom onset known
 Yes 205 84% 61 70% 104 90% 40 95%
 No 40 16% 26 30% 12 10% 2 5%
Time between symptom onset and PCR, daysa
 Median (IQR) 5 (2–7) 2 (1–7) 5 (3–8) 5 (3–7)
Time between disease onset and blood draw, daysb
 Median (IQR) 46 (24–62) 53 (42–66) 28 (18–59) 28 (26–41)
  1. aRefers to the first positive SARS-CoV-2 PCR of the patient. bRefers to the representative sample used in the sensitivity analysis. IQR, interquartile range; PCR, polymerase chain reaction.

In addition, 350 archived samples drawn in the pre-COVID-19 era were used to validate the specificity of the assays. In detail, 274 unselected samples dated from February 2017 to November 2019, 51 samples from hospitalized patients with bacterial pneumonia and 25 samples from patients with rheumatologic diseases were examined. Furthermore, an intravenous immunoglobulin formulation was tested (Supplementary Results).

Sample preparation

Blood samples were drawn according to routine clinical procedures. Upon centrifugation, serum specimens for antibody determination were kept at 4 °C if analyses were conducted within seven days or stored at −20 °C in the case analyses were conducted at a later time point. Frozen samples were thawed and centrifuged prior to antibody determination.

Anti-SARS-CoV-2 assays

We evaluated the performance of the following fully automated CLIA tests on high throughput random access analyzers widely available in medical laboratories: Roche Elecsys Anti-SARS-CoV-2 assay on the Cobas e602 platform (Roche Diagnostics, Rotkreuz, Switzerland), Abbott SARS-CoV-2 IgG assay on the Architect i2000SR platform (Abbott Laboratories Abbott Park, IL, USA), Siemens SARS-CoV-2 total (COV2T) and SARS-CoV-2 IgG (COV2G) on the Advia Centaur XP platform (Siemens, Munich, Germany). All samples were processed according to the manufacturers’ procedures with the specified controls and calibrators by trained laboratory staff. Test characteristics given in the manufacturers’ product information are summarized in Supplementary Table S1. Precision was evaluated by repeatedly measuring the positive controls of the respective assays (Supplementary Results).

Data analysis and statistics

Patients were included if at least one sample dating between day 14 and day 120 after disease onset was available. Only one sample per patient was subjected to sensitivity analysis, in case of multiple samples available per patient, the sample closest to day 28 was included.

Statistical analyses were performed using MedCalc, version 11.5.1.0 (MedCalc Ltd., Ostend, Belgium) and Excel 2016 (Microsoft, Redmont, USA). Median and interquartile range (IQR) were used as descriptive measures of metric data. Categorical data are given as counts and percentages. 95% confidence intervals (CI) for proportions were calculated according to the Clopper–Pearson exact method. The difference between categorical data was assessed using Chi-squared test (McNemar’s test for paired data, “N−1” Chi-squared test for unpaired proportions). Receiver operating characteristic (ROC) curve analysis was performed using the DeLong method. The concordance correlation coefficient was calculated according Lin. Statistical significance was defined at a level of 0.05.

Results

Specificity

Of the 350 pre-pandemic samples included, 341 were analyzed with the Roche Elecsys Anti-SARS-CoV-2 assay, 298 with Abbott SARS-CoV-2 IgG, 288 with Siemens SARS-CoV-2 total (COV2T), and 191 with Siemens SARS-CoV-2 IgG (COV2G). Using the manufacturers’ cut-offs, the specificity ranged from 99.3 to 100.0% (Roche 99.7%, Abbott 99.3%, Siemens COV2T 99.7%, Siemens COV2G 100.0%; Table 2). Simlar results were obtained when restricting the analysis to 146/350 (42%) samples tested with all assays (Supplementary Table S2). The small number of false positive samples (Supplementary Table S3) showed no overlap between the tests as none of the samples was tested positive in more than one assay. No false positive result was observed in the subcohorts of pre-pandemic samples from patients with bacterial pneumonia or rheumatologic diseases.

Table 2:

Sensitivity and specificity of the assays.

Investigated patient cohort Manufacturers’ claims
Roche Elecsys anti-SARS-CoV-2
 Sensitivity (n=230) 93.0% (89.0–96.0)a 99.5% (97.0–100.0)b
 Specificity (n=341) 99.7% (98.4–100.0) 99.8% (99.7–99.9)
Abbott SARS-CoV-2 IgG
 Sensitivity (n=228) 90.8% (86.3–93.9)a 100.0% (95.9–100.0)c
 Specificity (n=298) 99.3% (97.6–99.8) 99.6% (99.0–99.9)
Siemens SARS-CoV-2 total (COV2T)
 Sensitivity (n=195) 90.3% (85.3–93.7)a 100.0% (92.5–100.0)b
 Specificity (n=288) 99.7% (98.1–99.9) 99.8% (99.4–100.0)
Siemens SARS-CoV-2 IgG (COV2G)
 Sensitivity (n=193) 78.8% (72.5–83.9)a 100.0% (91.6–100.0)b
 Specificity (n=191) 100.0% (98.0–100.00) 99.9% (99.6–100.0)
  1. Sensitivity in the investigated cohort was evaluated using samples of SARS-CoV-2 PCR-confirmed patients dating ≥14 days after disease onset and the sample closest to day 28 after disease onset was chosen (one sample per patient). Specificity was determined on pre-pandemic samples. 95% confidence intervals are shown in brackets. a≥Day 14 after disease onset. b≥Day 14 after first PCR-positivity. c≥Day 14 after symptom onset.

Sensitivity

From 230 (93.9%) of the 245 PCR-confirmed COVID-19 patients, a sample dating between day 14 and day 120 after disease onset was available for analysis. The median time between disease onset and blood sampling was 46 days (IQR 24–62, range 14–120; Table 1; Supplementary Figure S1). Using the manufacturers’ cut-offs, the observed sensitivity of the assays ranged from 78.8 to 93.0% (Roche 93.0%, Abbott 90.8%, Siemens COV2T 90.3%, Siemens COV2G 78.8%; Table 2; Supplementary Table S4). When considering only the 189 (82%) samples which were tested with all four assays, sensitivities were virtually identical (Roche 92.6%, Abbott 90.5%, COV2T 90.5%, COV2G 79.4%, Supplementary Table S2). Compared to the COV2G assay, the Roche (p<0.0001), Abbott (p=0.0001) and COV2T (p<0.0001) tests were significantly more sensitive.

In an exploratory analysis, we assessed the sensitivity of the assays stratified for gender, age, and disease severity (Supplementary Table S4). When stratifying for gender, sensitivity was higher for males than for females within the Abbott (male: 94.5%, female: 84.2%; p=0.0095) and Siemens COV2G (male: 87.3%, female: 62.7%; p=0.0001) assays (Figure 1A). Stratified for patient age, sensitivity for Siemens COV2G was lower in the age group 18–49 years (63.9%) compared to patients aged 50–69 years (86.3%) and 70–100 years (83.8%) (Figure 1B). Next, we stratified the patients according to intensity of care needed as a surrogate for disease severity. For outpatients sensitivity ranged from 64.9 to 91.8%, for hospitalized patients at the general ward from 84.3 to 93.4%, and for ICU patients from 93.9 to 96.7%. In particular, the COV2G test had a significantly lower sensitivity in outpatient compared to hospitalized patients (Figure 1C).

Figure 1: 
Subgroup analyses of sensitivity according to gender, age and severity of disease.
Comparison of the sensitivity of the four investigated tests in representative samples of 230 patients with PCR-confirmed COVID-19: Roche (Roche Elecsys anti-SARS-CoV-2, blue), Abbott (Abbott SARS-CoV-2 IgG, green), Siemens COV2T (Siemens SARS-CoV-2 total, orange) and Siemens COVG (Siemens SARS-CoV-2 IgG, brown). Results were analyzed stratified for gender (A), age (B) and severity of disease (C).
Figure 1:

Subgroup analyses of sensitivity according to gender, age and severity of disease.

Comparison of the sensitivity of the four investigated tests in representative samples of 230 patients with PCR-confirmed COVID-19: Roche (Roche Elecsys anti-SARS-CoV-2, blue), Abbott (Abbott SARS-CoV-2 IgG, green), Siemens COV2T (Siemens SARS-CoV-2 total, orange) and Siemens COVG (Siemens SARS-CoV-2 IgG, brown). Results were analyzed stratified for gender (A), age (B) and severity of disease (C).

To address potential reasons for unexpected negative results, clinical records of patients tested negative with the majority of SARS-CoV-2 antibody tests were analyzed in detail. Out of the 230 patient samples included in the primary endpoint analysis, 13 samples gave negative results in all performed tests and two samples in all but one assay. Seven of those patients were under immunosuppressive treatment not related to COVID-19 (cancer with ongoing chemotherapy n=3, lymphoma with anti-CD20 immunotherapy n=2, rheumatoid arthritis with methotrexate n=1, and myasthenia gravis under treatment with corticosteroids and rituximab n=1). The other eight patients did not have relevant immunosuppressive conditions or therapies and total serum IgG, IgM and IgA immunoglobulins were within the reference range.

Positivity rate across the course of disease

To estimate the time to positivity after disease onset, we analyzed all 700 samples from 245 PCR-confirmed COVID-19 patients and stratified the positivity rate for the time from disease onset. For all tests, the positivity rate increased from week one (range 19.5–32.1%) to week two (range 40.6–54.6%) and week three (range 80.9–95.3%). While the positivity rate after week three slightly decreased for Roche and Abbott, the positivity rate of samples for Siemens COV2T and Siemens COV2G increased from day 14–20 to day 21–40 and dropped again after day 40 (Figure 2; Supplementary Table S5, Figure S2). The latter effect might be confounded by the higher number of outpatient samples in the group of samples obtained >40 days from disease onset (Supplementary Figure S1).

Figure 2: 
Positivity rate across the course of disease. All 700 samples from 245 PCR-confirmed SARS-CoV-2 infected patients were subjected to antibody determination with Roche (Roche Elecsys anti-SARS-CoV-2, n=695, blue), Abbott (Abbott SARS-CoV-2 IgG, n=684, green), Siemens COV2T (Siemens SARS-CoV-2 total, n=517, orange) and Siemens COVG (Siemens SARS-CoV-2 IgG, n=498, brown). The positivity rate of the respective tests is shown in dependence of the time between disease onset and blood draw in days (d).
Figure 2:

Positivity rate across the course of disease. All 700 samples from 245 PCR-confirmed SARS-CoV-2 infected patients were subjected to antibody determination with Roche (Roche Elecsys anti-SARS-CoV-2, n=695, blue), Abbott (Abbott SARS-CoV-2 IgG, n=684, green), Siemens COV2T (Siemens SARS-CoV-2 total, n=517, orange) and Siemens COVG (Siemens SARS-CoV-2 IgG, n=498, brown). The positivity rate of the respective tests is shown in dependence of the time between disease onset and blood draw in days (d).

ROC curve analysis

ROC curve analysis revealed area under the curve (AUC) values of 0.984 for Roche, 0.982 for Abbott, 0.975 for Siemens COV2T and 0.966 for Siemens COV2G (Figure 3; Supplementary Table S6). When comparing the ROC curves, the AUC of the COV2G test was significantly smaller compared to Roche (p=0.0151) and Abbott (p=0.0174). In an exploratory analysis, we asked if the assay cut-offs could be modified to improve the sensitivity. The proposed cut-off indexes (COI) based on the maximum sum of sensitivity and specificity in ROC curve analysis were below the manufacturers’ COI: >0.15 for Roche, >0.54 for Abbott, >0.42 for Siemens COV2T and >0.32 for Siemens COV2G (Figure 3). Using these modified COI instead of the manufacturers’ COI, sensitivity improved from 93.0 to 95.7% for Roche, from 90.8 to 95.2% for Abbott, from 90.3 to 92.3% for Siemens COV2T and from 78.8 to 89.6% for Siemens COV2G. However, specificity of the assays diminished accordingly (Supplementary Table S7).

Figure 3: 
ROC curve analysis and modified cut-off indexes.
ROC curves (left) and corresponding raw data indexes (right) for the investigated tests from Roche (Roche Elecsys anti-SARS-CoV-2, A and B, blue), Abbott (Abbott SARS-CoV-2 IgG, C and D, green), Siemens COV2T (Siemens SARS-CoV-2 total, E and F, orange) and Siemens COV2G (Siemens SARS-CoV-2 IgG, G and H, brown). Analyses are based on pre-pandemic (=negative) samples (samples, Roche: 341, Abbott: 298, Siemens COV2T: 288, Siemens COV2G: 191) and all the samples of SARS-CoV-2 PCR-confirmed patients dating ≥14 days after disease onset (samples, Roche: 537, Abbott: 530, Siemens COV2T: 392, Siemens COV2G: 375). The horizontal lines in the dot plots represent the manufacturers’ cut off index (dashed line, Roche and Siemens: 1.0, Abbot: 1.4) and the modified cut off index according to ROC curve analysis (solid line).
Figure 3:

ROC curve analysis and modified cut-off indexes.

ROC curves (left) and corresponding raw data indexes (right) for the investigated tests from Roche (Roche Elecsys anti-SARS-CoV-2, A and B, blue), Abbott (Abbott SARS-CoV-2 IgG, C and D, green), Siemens COV2T (Siemens SARS-CoV-2 total, E and F, orange) and Siemens COV2G (Siemens SARS-CoV-2 IgG, G and H, brown). Analyses are based on pre-pandemic (=negative) samples (samples, Roche: 341, Abbott: 298, Siemens COV2T: 288, Siemens COV2G: 191) and all the samples of SARS-CoV-2 PCR-confirmed patients dating ≥14 days after disease onset (samples, Roche: 537, Abbott: 530, Siemens COV2T: 392, Siemens COV2G: 375). The horizontal lines in the dot plots represent the manufacturers’ cut off index (dashed line, Roche and Siemens: 1.0, Abbot: 1.4) and the modified cut off index according to ROC curve analysis (solid line).

Concordance between results of different antibody assays indicates a potential for test combinations

When considering all 1,050 samples the concordance correlation coefficient between the four assays ranged from 77.5 to 92.5% (Supplementary Table S8). Of the 230 available COVID-19 patient samples, 189 samples were tested with all the four assays. 145 (76.7%) of these were positive with all four assays, and 11 (5.8%) were found negative in all four assays. Of note, 18 samples (9.5%) were negative only with COV2G (Figure 4). Due to the unexpected low sensitivity of the Siemens COV2G test, we restricted further analyses on combinatorial approaches to the three assays Roche, Abbott and Siemens COV2T. Of the 350 pre-pandemic samples included in our study, 235 were tested with these three assays, and 233 of these samples (99.1%) were concordantly negative.

Figure 4: 
Concordance of the SARS-CoV-2 antibody results in PCR-confirmed patients.
Overlap of negative (A) and positive (B) antibody test results of different assays in SARS-CoV-2 PCR-confirmed COVID-19 patients (samples dating ≥14 days after disease onset; one representative sample per patient; a total of 189 samples in which data from all four assays were available). Abbott, Abbott SARS-CoV-2 IgG; Roche, Roche Elecsys anti-SARS-CoV-2; COV2G, Siemens SARS-CoV-2 IgG; COV2T, Siemens SARS-CoV-2 total.
Figure 4:

Concordance of the SARS-CoV-2 antibody results in PCR-confirmed patients.

Overlap of negative (A) and positive (B) antibody test results of different assays in SARS-CoV-2 PCR-confirmed COVID-19 patients (samples dating ≥14 days after disease onset; one representative sample per patient; a total of 189 samples in which data from all four assays were available). Abbott, Abbott SARS-CoV-2 IgG; Roche, Roche Elecsys anti-SARS-CoV-2; COV2G, Siemens SARS-CoV-2 IgG; COV2T, Siemens SARS-CoV-2 total.

Finally, we asked whether a combination of two of these antibody tests could be useful to further improve the clinical performance of SARS-CoV-2 serology. When combining any two out of the Roche, Abbott and Siemens COV2T assays and when defining as positive only samples with concordantly positive results in both tests, specificity improves to 100.0%, regardless of the combination used. However, sensitivity dropped to <90% (Table 3). When we used the modified COI instead of the manufacturers’ COI, sensitivity of the combinations improved to 95.2% for Roche and Abbott (manufacturers’ COI: 89.9%), 92.3% for Roche and Siemens COV2T (manufacturers’ COI: 89.2%), and 92.3% for Abbott and Siemens COV2T (manufacturers’ COI: 86.6%). In contrast, the specificity of the combined analyses was not affected by the lowering of the COI (Table 3).

Table 3:

Sensitivity and specificity for the combination of two tests.

Manufacturers’ COI Modified COI
Roche + Abbott
 Combined sensitivity (n=228) 89.9% (85.3–93.2%) 95.2% (91.6–97.3%)
 Combined specificity (n=291) 100.0% (98.7–100.0%) 100.0% (98.7–100.0%)
Roche + Siemens COV2T
 Combined sensitivity (n=195) 89.2% (84.1–92.9%) 92.3% (87.7–95.3%)
 Combined specificity (n=282) 100.0% (98.7–100.0%) 100.0% (98.7–100.0%)
Abbott + Siemens COV2T
 Combined sensitivity (n=194) 86.6% (81.1–90.7%) 92.3% (87.6–95.3%)
 Combined specificity (n=241) 100.0% (98.4–100.0%) 100.0% (98.4–100.0%)
  1. Sensitivity and specificity for the combination of assays (double positive samples were regarded as positive, single or double negative samples were regarded as negative) using different cut-off indexes (COI): the COI according to the manufacturer and a modified COI according to ROC curve analysis of the results of our cohort. Sensitivity in the investigated cohort was evaluated using samples of SARS-CoV-2 PCR-confirmed patients dating ≥14 days after disease onset and the sample closest to day 28 after disease onset was chosen (one sample per patient). Specificity was determined on pre-pandemic samples. 95% confidence intervals are shown in brackets. Roche, Roche Elecsys anti-SARS-CoV-2; Abbott, Abbott SARS-CoV-2 IgG; Siemens COV2T, Siemens SARS-CoV-2 total.

Discussion

In the present study we examined the performance characteristics of four fully automated SARS-CoV-2 antibody CLIA assays focusing on high throughput random access analyzers widely available in many medical laboratories. While the specificity of all assays was well comparable to the performance characteristics provided by the manufacturers, we observed a markedly lower sensitivity in our cohort (78.8–93.0% compared to >99%). Similar results have been found in other studies for the Roche, Abbott and the Siemens COV2T assays [27], [28]. These findings emphasize the importance of real life data and different clinical scenarios in independent assay validations. While the more subtle differences between observed and expected sensitivity rates for Roche, Abbott and Siemens COV2T were coherent within our study and might be explained by specific characteristics of our patients cohort, the sensitivity of Siemens COV2G was markedly lower in our hands with 78.8% in the total cohort and 64.9% in outpatients. To the best of our knowledge, our study is the first public-available independent evaluation of the COV2G assay and warrants further studies to verify and potentially improve the performance of this test. Indeed, the company has just recently filed an FDA-application for approval of a new SARS-CoV-2 IgG assay (sCOVG) with improved sensitivity. Our results indicate that potential differences in the assay performance between the Siemens COV2G and the new sCOVG assays need to be considered when evaluating sequential samples of patients.

Interestingly, we found lower sensitivities in female vs. male patients with all four assays, whereby a statistical significant difference was found only in the Abbott and Siemens COV2G assays. In a sub-analysis this trend for a gender difference was found independent of the courses of disease. This is grossly in line with the findings of Korte et al. [29], which showed that men produce higher amounts of anti SARS-CoV-2 IgG and IgA. Inconsistent results have been reported regarding the association between antibody titers and disease severity. For example, in a serosurvey in health care workers of the Veneto Region of Italy, Plebani et al. [30] found that symptomatic individuals were 100% SARS-CoV-2 antibody positive, whilst only in 58% of asymptomatic carriers antibodies were detectable. Phipps et al. [31] could not find an association of IgG and IgM antibody response and disease severity, however, patients were segregated according to need of ICU and no asymptomatic or outpatient population was described separately. The importance of including patients with mild disease course in antibody evaluation studies has already been highlighted [24]. We only observed significantly lower positivity rates in outpatients for the Siemens COV2G test but not for the Roche, Abbott and the Siemens COV2T assays. However, quantitative differences in antibody levels were not assessed in our study.

Notably, absence of reactivity in serological assays could either reflect antibody test performance or the biological absence of antibodies in individuals [32]. As 5.7% of patients were concordantly negative in all assays and comparable frequencies were found by other authors in different cohorts (Baron et al. 3% [33], Petersen et al. 6% [34], Oved et al. 5% [35]), we conclude that the biological absence of antibodies was a relevant factor in our cohort. This might be explained by immediate infection clearance in the naso-pharyngeal space as a consequence of low viral exposure and/or effective immune function which does not induce a systemic immune response or by immunosuppression. While no data on T-cell mediated immune response were available in our patients, detailed clinical meta-data allowed to correlate the absence of humoral immune response with immunosuppression. Indeed, a number of these patients was immunocompromised what likely explains the absence of a detectable systemic humoral immune response, and warrants for careful interpretation of SARS-CoV-2 antibody results in those.

SARS-CoV-2 seroprevalence is still used as a measure to estimate the true number of affected people during the pandemic. Such studies found seropositivity for SARS-CoV-2 antibodies in 2.4% (n=61,437) of residents in Wuhan in mid-May [36], 0.9% (n=3,186) in German regular blood donors from March to June [37], 1.0% (n=2,500) in Greece university personnel and students during June-July [38], and 4.6% (n=5,933) of health care workers till end of May at the University-Hospitals of Padova and Verona, Italy [30]. Thus, a very high diagnostic specificity of serologic tests is crucial to minimize false positive results and improve the positive predictive value (PPV). In line with the performance characteristics provided by the manufacturers, we observed a false positive rate <1% for all tests using samples from the pre-pandemic era. However, even a small false positive rate substantially effects the PPV in a low-prevalence situation [12]. It has thus been suggested to combine the results of two different SARS-CoV-2 antibody tests to further improve specificity and PPV of serologic testing [39], [40], [41]. As a prerequisite for this strategy, it has been shown that the vast majority of false positive results occurred independently with singular anti-SARS-CoV-2 CLIA assays [12]. This is also in line with our observation. In contrast, the concordance of false negative samples between the Roche, Abbott and Siemens COV2T assays was rather high and yielded in a combined sensitivity of >86% for any of these combinations. While singular tests have been optimized for maximal specificity, a combinatorial testing approach could allow to lower the thresholds to recognize borderline reactive samples without impairing specificity [40]. Lower cut-offs similar to our findings of the ROC analysis have already been described [42]. With these lower cut-offs, the combined sensitivity improved to >92% for any combination. These results suggest, that borderline reactive results slightly below the threshold of the test could benefit from further analyses with additional CLIA or ELISA assays. However, due to the huge impact of the disease prevalence and pretest probability on predictive values of the assay results, the purpose of the test (detection or exclusion of disease, seroprevalence studies) should be defined before adjustment of thresholds or combining different serologic tests in a certain setting [43]. Likewise, orthogonal testing algorithms have been proposed by the CDC [44], [45]. Furthermore, a multiplex immunoassay detecting simultaneously the presence of antibodies directed against diverse antigens (N, S1, S2, RBD, NTD) in the same test was shown to improve the overall specificity and sensitivity [46].

Limitations of our study include differences in the time between disease onset and serological assessment between patients due to the retrospective design of this assay validation study. To avoid potential biases, we constrained our analysis to samples drawn ≥14 days after onset of COVID-19 specific symptoms in SARS-CoV-2 RT-PCR confirmed patients or two weeks after the first positive RT-PCR result in patients for whom no information on symptom onset was available. The median time of 46 days between disease onset and date of the sample used for the sensitivity analysis fits well to the reported plateau of antibody production against SARS-CoV-2 [47]. Thus, our study was designed to assess the maximal sensitivity of antibody tests and did not include the early phase of antibody production within the first days of COVID-19. Likewise, the maximum time between disease onset and sampling was 120 days. The duration of antibody production and immunity after a SARS-CoV-2 infection is a major question. While some studies described a significant loss of seropositivity during long time follow-up [48], [49], others did not [50]. However, among others, the observed discrepancies may also be due to differences in the serologic assays (e.g. pan-immunoglobulin vs. immunoglobulin isotypes or different target antigens) used in the different studies [51], [52]. In our study, no substantial decrease of seropositivity was found within 120 days after disease onset. Further studies are needed to assess the performance of different serological tests in longitudinal analysis. In this regard, quantitative measurements of antibody titers and neutralization or pseudo-neutralization assays will be useful to monitor the course of the humoral immune response against SARS-CoV-2 in detail. Another limitation of our study was that only adult patients with COVID-19 have been included. For example, Pierce et al. found that serum neutralizing antibody titers were higher in adults compared to pediatric patients with COVID-19 [53]. Additional studies are needed to evaluate the performance of the assays tested here for children.

In summary, we independently evaluated the performance of four widely available commercial SARS-CoV-2 antibody assays in an adult COVID-19 cohort including both patients with critical to severe as well as mild courses of the disease. While all assays met the desired specificity criteria, we observed a substantially lower sensitivity compared to the performance reported by the manufacturers. Our study emphasizes the importance of achieving additional performance data in real life including specific populations like immunocompromised patients and asymptomatic carriers. Importantly, our results suggest a limited sensitivity of the Siemens COV2G assay that will be replaced by the newly filed Siemens sCOVG assay. In conclusion, a growing number of fully automated SARS-CoV-2 antibody CLIA assays with sufficient performance characteristics is available for different high throughput analyzers. The selection of specific assays and the interpretation of results should carefully reflect the use case, and a combination of different SARS-CoV-2 antibody assays might be useful.


Corresponding author: Gregor Hoermann, MD, PhD, Central Institute of Clinical and Chemical Laboratory Diagnostics, University Hospital Innsbruck, Anichstrasse 35A-6020, Innsbruck, Austria; and MLL Munich Leukemia Laboratory, Munich, Germany, Phone: +43 50 504 24091, Fax: +43 50 504 24088, E-mail:

Acknowledgments

The authors thank Bettina Schatz, Daniel Auβerdorfer (both Central Institute of Clinical and Chemical Laboratory Diagnostics, University Hospital Innsbruck) and Gernot Osterer (Siemens) for technical support and are grateful to Manfred Herold (Department of Internal Medicine II, University Hospital Innsbruck) for providing serum samples of patients with rheumatologic diseases.

  1. Research funding: The study was performed by institutional research funding.

  2. Author contributions: CI, AE, WP, LL, MA, AG, and GH performed or analyzed serological tests. SS, AP, TS, WM, HS, JL-R, RB-W, IT, and GW obtained clinical data. CI and AE performed statistical analysis. CI, AE, AG, and GH designed the study and wrote the paper. All authors revised and approved the manuscript.

  3. Competing interests: IT was awarded an Investigator-Initiated Study (IIS) grant by Boehringer Ingelheim (IIS 1199-0424). All other authors state no conflict of interest.

  4. Ethical approval: All procedures performed in the present study involving human participants were in accordance with the ethical standards of the Institutional and/or National Research Committee and with the 1964 Helsinki declaration and its later amendments and were approved by the Ethics Committee of the Medical University of Innsbruck (Ethics Commission numbers: 1103/2020, 1167/2020).

References

1. Zhu, N, Zhang, D, Wang, W, Li, X, Yang, B, Song, J, et al.. A novel coronavirus from patients with pneumonia in China, 2019. N Engl J Med 2020;382:727–33.10.1056/NEJMoa2001017Search in Google Scholar

2. Guan, WJ, Ni, ZY, Hu, Y, Liang, WH, Ou, CQ, He, JX, et al.. Clinical characteristics of coronavirus disease 2019 in China. N Engl J Med 2020;382:1708–20.10.1056/NEJMoa2002032Search in Google Scholar

3. World Health Organization. Virtual press conference on COVID-19 – 11 March 2020. Available from: https://www.who.int/docs/default-source/coronaviruse/transcripts/who-audio-emergencies-coronavirus-press-conference-full-and-final-11mar2020.pdf?sfvrsn=cb432bb3_2 [Accessed 20 Oct 2020].Search in Google Scholar

4. Center for Systems Science and Engineering (CSSE) at Johns Hopkins University. COVID-19 dashboard. Available from: https://coronavirus.jhu.edu/map.html [Accessed 24 Nov 2020].Search in Google Scholar

5. Nishiura, H, Kobayashi, T, Yang, Y, Hayashi, K, Miyama, T, Kinoshita, R, et al.. The rate of underascertainment of novel coronavirus (2019-nCoV) infection: Estimation using Japanese passengers data on Evacuation flights. J Clin Med 2020;9:419.10.3390/jcm9020419Search in Google Scholar

6. Ravi, N, Cortade, DL, Ng, E, Wang, SX. Diagnostics for SARS-CoV-2 detection: a comprehensive review of the FDA-EUA COVID-19 testing landscape. Biosens Bioelectron 2020;165:112454.10.1016/j.bios.2020.112454Search in Google Scholar

7. Li, C, Zhao, C, Bao, J, Tang, B, Wang, Y, Gu, B. Laboratory diagnosis of coronavirus disease-2019 (COVID-19). Clin Chim Acta 2020;510:35–46.10.1016/j.cca.2020.06.045Search in Google Scholar

8. To, KK, Tsang, OT, Leung, WS, Tam, AR, Wu, TC, Lung, DC, et al.. Temporal profiles of viral load in posterior oropharyngeal saliva samples and serum antibody responses during infection by SARS-CoV-2: an observational cohort study. Lancet Infect Dis 2020;20:565–74.10.1016/S1473-3099(20)30196-1Search in Google Scholar

9. Young, BE, Ong, SWX, Kalimuddin, S, Low, JG, Tan, SY, Loh, J, et al.. Epidemiologic features and clinical course of patients infected with SARS-CoV-2 in Singapore. J Am Med Assoc 2020;323:1488–94.10.1001/jama.2020.3204Search in Google Scholar PubMed PubMed Central

10. Hu, Z, Song, C, Xu, C, Jin, G, Chen, Y, Xu, X, et al.. Clinical characteristics of 24 asymptomatic infections with COVID-19 screened among close contacts in Nanjing, China. Sci China Life Sci 2020;63:706–11.10.1007/s11427-020-1661-4Search in Google Scholar PubMed PubMed Central

11. Azkur, AK, Akdis, M, Azkur, D, Sokolowska, M, van de Veen, W, Bruggen, MC, et al.. Immune response to SARS-CoV-2 and mechanisms of immunopathological changes in COVID-19. Allergy 2020;75:1564–81.10.1111/all.14364Search in Google Scholar PubMed PubMed Central

12. Perkmann, T, Perkmann-Nagele, N, Breyer, MK, Breyer-Kohansal, R, Burghuber, OC, Hartl, S, et al.. Side-by-Side comparison of three fully automated SARS-CoV-2 antibody assays with a focus on specificity. Clin Chem 2020;66:1405–13.10.1093/clinchem/hvaa198Search in Google Scholar PubMed PubMed Central

13. Zhao, J, Yuan, Q, Wang, H, Liu, W, Liao, X, Su, Y, et al.. Antibody responses to SARS-CoV-2 in patients of novel coronavirus disease 2019. Clin Infect Dis 2020;71:2027–34.10.1093/cid/ciaa344Search in Google Scholar PubMed PubMed Central

14. European Commission. Directorate-general for health and food safety. An EU programme of COVID-19 convalescent plasma collection and transfusion. Guidance on collection, testing, processing, storage, distribution and monitored use. 2020.Search in Google Scholar

15. Jackson, LA, Anderson, EJ, Rouphael, NG, Roberts, PC, Makhene, M, Coler, RN, et al.. An mRNA vaccine against SARS-CoV-2 - preliminary report. N Engl J Med 2020;383:1920–31.10.1056/NEJMoa2022483Search in Google Scholar PubMed PubMed Central

16. Bohn, MK, Lippi, G, Horvath, A, Sethi, S, Koch, D, Ferrari, M, et al.. Molecular, serological, and biochemical diagnosis and monitoring of COVID-19: IFCC taskforce evaluation of the latest evidence. Clin Chem Lab Med 2020;58:1037–52.10.1515/cclm-2020-0722Search in Google Scholar PubMed

17. Braun, J, Loyal, L, Frentsch, M, Wendisch, D, Georg, P, Kurth, F, et al.. SARS-CoV-2-reactive T cells in healthy donors and patients with COVID-19. Nature 2020;587:270–4.10.1038/s41586-020-2598-9Search in Google Scholar PubMed

18. Vashist, SK. In vitro diagnostic assays for COVID-19: recent advances and emerging trends. Diagnostics 2020;10:202.10.3390/diagnostics10040202Search in Google Scholar PubMed PubMed Central

19. Espejo, AP, Akgun, Y, Al Mana, AF, Tjendra, Y, Millan, NC, Gomez-Fernandez, C, et al.. Review of current advances in serologic testing for COVID-19. Am J Clin Pathol 2020;154:293–304.10.1093/ajcp/aqaa112Search in Google Scholar PubMed PubMed Central

20. Poh, CM, Carissimo, G, Wang, B, Amrun, SN, Lee, CY, Chee, RS, et al.. Two linear epitopes on the SARS-CoV-2 spike protein that elicit neutralising antibodies in COVID-19 patients. Nat Commun 2020;11:2806.10.1038/s41467-020-16638-2Search in Google Scholar PubMed PubMed Central

21. Theel, ES, Harring, J, Hilgart, H, Granger, D. Performance characteristics of four high-throughput immunoassays for detection of IgG antibodies against SARS-CoV-2. J Clin Microbiol 2020;58:e01243-20. https://doi.org/10.1128/jcm.01243-20.10.1128/JCM.01243-20Search in Google Scholar PubMed PubMed Central

22. Lisboa Bastos, M, Tavaziva, G, Abidi, SK, Campbell, JR, Haraoui, LP, Johnston, JC, et al.. Diagnostic accuracy of serological tests for COVID-19: systematic review and meta-analysis. BMJ 2020;370:m2516.10.1136/bmj.m2516Search in Google Scholar PubMed PubMed Central

23. Deeks, JJ, Dinnes, J, Takwoingi, Y, Davenport, C, Spijker, R, Taylor-Phillips, S, et al.. Antibody tests for identification of current and past infection with SARS-CoV-2. Cochrane Database Syst Rev 2020;6:CD013652.10.1002/14651858.CD013652Search in Google Scholar PubMed PubMed Central

24. Patel, E, Bloch, EM, Clarke, W, Hsieh, YH, Boon, D, Eby, YJ, et al.. Comparative performance of five commercially available serologic assays to detect antibodies to SARS-CoV-2 and identify individuals with high neutralizing titers. medRxiv 2020.10.1101/2020.08.31.20184788Search in Google Scholar

25. Plebani, M. Antibody responses in mild COVID-19 hospital staff. EBioMedicine 2020;59:102940.10.1016/j.ebiom.2020.102940Search in Google Scholar

26. WHO Working Group on the Clinical Characterisation and Management of Covid-19 infection. A minimal common outcome measure set for COVID-19 clinical research. Lancet Infect Dis 2020;20:e192–7.10.1016/S1473-3099(20)30483-7Search in Google Scholar

27. Schnurra, C, Reiners, N, Biemann, R, Kaiser, T, Trawinski, H, Jassoy, C. Comparison of the diagnostic sensitivity of SARS-CoV-2 nucleoprotein and glycoprotein-based antibody tests. J Clin Virol 2020;129:104544.10.1016/j.jcv.2020.104544Search in Google Scholar

28. Manthei, DM, Whalen, JF, Schroeder, LF, Sinay, AM, Li, SH, Valdez, R, et al.. Differences in performance characteristics among four high-throughput assays for the detection of antibodies against SARS-CoV-2 using a common set of patient samples. Am J Clin Pathol 2020;XX:1–13. https://doi.org/10.1093/ajcp/aqaa200.10.1093/ajcp/aqaa200Search in Google Scholar

29. Korte, W, Buljan, M, Rosslein, M, Wick, P, Golubov, V, Jentsch, J, et al.. SARS-CoV-2 IgG and IgA antibody response is gender dependent; and IgG antibodies rapidly decline early on. J Infect 2020. https://doi.org/10.1016/j.jinf.2020.08.032.10.1016/j.jinf.2020.08.032Search in Google Scholar

30. Plebani, M, Padoan, A, Fedeli, U, Schievano, E, Vecchiato, E, Lippi, G, et al.. SARS-CoV-2 serosurvey in health care workers of the Veneto Region. Clin Chem Lab Med 2020;58:2107–11.10.1515/cclm-2020-1236Search in Google Scholar

31. Phipps, WS, SoRelle, JA, Li, QZ, Mahimainathan, L, Araj, E, Markantonis, J, et al.. SARS-CoV-2 antibody responses do not predict COVID-19 disease severity. Am J Clin Pathol 2020;154:459–65.10.1093/ajcp/aqaa123Search in Google Scholar

32. National, S-C-SAEG. Performance characteristics of five immunoassays for SARS-CoV-2: a head-to-head benchmark comparison. Lancet Infect Dis 2020;20:1390–400.10.1016/S1473-3099(20)30634-4Search in Google Scholar

33. Baron, RC, Risch, L, Weber, M, Thiel, S, Grossmann, K, Wohlwend, N, et al.. Frequency of serological non-responders and false-negative RT-PCR results in SARS-CoV-2 testing: a population-based study. Clin Chem Lab Med 2020;58:2131–40.10.1515/cclm-2020-0978Search in Google Scholar PubMed

34. Petersen, LR, Sami, S, Vuong, N, Pathela, P, Weiss, D, Morgenthau, BM, et al.. Lack of antibodies to SARS-CoV-2 in a large cohort of previously infected persons. Clin Infect Dis 2020;ciaa1685. https://doi.org/10.1093/cid/ciaa1685.10.1093/cid/ciaa1685Search in Google Scholar PubMed PubMed Central

35. Oved, K, Olmer, L, Shemer-Avni, Y, Wolf, T, Supino-Rosin, L, Prajgrod, G, et al.. Multi-center nationwide comparison of seven serology assays reveals a SARS-CoV-2 non-responding seronegative subpopulation. EClinicalMedicine 2020;29:100651. https://doi.org/10.1016/j.eclinm.2020.100651.10.1016/j.eclinm.2020.100651Search in Google Scholar PubMed PubMed Central

36. Pan, Y, Li, X, Yang, G, Fan, J, Tang, Y, Hong, X, et al.. Seroprevalence of SARS-CoV-2 immunoglobulin antibodies in Wuhan, China: part of the city-wide massive testing campaign. Clin Microbiol Infect 2020. https://doi.org/10.1016/j.cmi.2020.09.044.10.1016/j.cmi.2020.09.044Search in Google Scholar PubMed PubMed Central

37. Fischer, B, Knabbe, C, Vollmer, T. SARS-CoV-2 IgG seroprevalence in blood donors located in three different federal states, Germany, March to June 2020. Euro Surveill 2020;25:2001285.10.2807/1560-7917.ES.2020.25.28.2001285Search in Google Scholar PubMed PubMed Central

38. Tsitsilonis, OE, Paraskevis, D, Lianidou, E, Pierros, V, Akalestos, A, Kastritis, E, et al.. Seroprevalence of antibodies against SARS-CoV-2 among the personnel and students of the National and Kapodistrian university of athens, Greece: a preliminary report. Life 2020;10:214.10.3390/life10090214Search in Google Scholar PubMed PubMed Central

39. Suhandynata, RT, Hoffman, MA, Kelner, MJ, McLawhon, RW, Reed, SL, Fitzgerald, RL. Multi-platform comparison of SARS-CoV-2 serology assays for the detection of COVID-19. J Appl Lab Med 2020;5:1324–36.10.1093/jalm/jfaa139Search in Google Scholar PubMed PubMed Central

40. Perkmann, TP-NN, Oszvar-Kozma, M, Koller, T, Breyer, M-K, Breyer-Kohansal, R, et al.. Increasing both specificity and sensitivity of SARS-CoV-2 antibody tests by using an adaptive orthogonal testing approach. medRxiv 2020.Search in Google Scholar

41. Risch, M, Weber, M, Thiel, S, Grossmann, K, Wohlwend, N, Lung, T, et al.. Temporal course of SARS-CoV-2 antibody positivity in patients with COVID-19 following the first clinical presentation. BioMed Res Int 2020;2020:9878453.10.1155/2020/9878453Search in Google Scholar PubMed PubMed Central

42. Favresse, J, Eucher, C, Elsen, M, Tre-Hardy, M, Dogne, JM, Douxfils, J. Clinical performance of the Elecsys Electrochemiluminescent immunoassay for the detection of SARS-CoV-2 total antibodies. Clin Chem 2020;66:1104–6. https://doi.org/10.1093/clinchem/hvaa131.Search in Google Scholar

43. Plebani, M, Padoan, A, Sciacovelli, L, Basso, D. Towards the rational utilization of SARS-CoV-2 serological tests in clinical practice. Clin Chem Lab Med 2020;58:e189–91.10.1515/cclm-2020-0880Search in Google Scholar PubMed

44. Centers for Disease Control and Prevention. Interim guidelines for COVID-19 antibody testing in clinical and public health settings. Available from: https://www.cdc.gov/coronavirus/2019-ncov/lab/resources/antibody-tests-guidelines.html [Accessed 18 Dec 2020].10.46234/ccdcw2020.085Search in Google Scholar PubMed PubMed Central

45. Food and Drug Administration. Calculator for positive predictive value (PPV) and negative predictive value (NPV) for individual tests and combined. Available from: https://www.fda.gov/media/137612/download [Accessed 18 Dec 2020].Search in Google Scholar

46. Gillot, C, Douxfils, J, Cadrobbi, J, Laffineur, K, Dogne, JM, Elsen, M, et al.. An original ELISA-based multiplex method for the simultaneous detection of 5 SARS-CoV-2 IgG antibodies directed against different antigens. J Clin Med 2020;9:3752.10.3390/jcm9113752Search in Google Scholar PubMed PubMed Central

47. Favresse, J, Eucher, C, Elsen, M, Laffineur, K, Dogne, JM, Douxfils, J. Response of anti-SARS-CoV-2 total antibodies to nucleocapsid antigen in COVID-19 patients: a longitudinal study. Clin Chem Lab Med 2020;58:e193–6. https://doi.org/10.1515/cclm-2020-0962.Search in Google Scholar

48. Long, QX, Tang, XJ, Shi, QL, Li, Q, Deng, HJ, Yuan, J, et al.. Clinical and immunological assessment of asymptomatic SARS-CoV-2 infections. Nat Med 2020;26:1200–4. https://doi.org/10.1038/s41591-020-0965-6.Search in Google Scholar

49. Liu, A, Li, Y, Peng, J, Huang, Y, Xu, D. Antibody responses against SARS-CoV-2 in COVID-19 patients. J Med Virol 2020. https://doi.org/10.1002/jmv.26646.Search in Google Scholar

50. Isho, B, Abe, KT, Zuo, M, Jamal, AJ, Rathod, B, Wang, JH, et al.. Persistence of serum and saliva antibody responses to SARS-CoV-2 spike antigens in COVID-19 patients. Sci Immunol 2020;5:eabe5511. https://doi.org/10.1126/sciimmunol.abe5511.10.1126/sciimmunol.abe5511Search in Google Scholar PubMed PubMed Central

51. Schaffner, A, Risch, L, Weber, M, Thiel, S, Jüngert, K, Pichler, M, et al.. Sustained SARS-CoV-2 nucleocapsid antibody levels in nonsevere COVID-19: a population-based study. Clin Chem Lab Med 2021;59:e49–51.10.1515/cclm-2020-1347Search in Google Scholar PubMed

52. Gudbjartsson, DF, Norddahl, GL, Melsted, P, Gunnarsdottir, K, Holm, H, Eythorsson, E, et al.. Humoral immune response to SARS-CoV-2 in Iceland. N Engl J Med 2020;383:1724–34.10.1056/NEJMoa2026116Search in Google Scholar PubMed PubMed Central

53. Pierce, CA, Preston-Hurlburt, P, Dai, Y, Aschner, CB, Cheshenko, N, Galen, B, et al.. Immune responses to SARS-CoV-2 infection in hospitalized pediatric and adult patients. Sci Transl Med 2020;12:eabd5487. https://doi.org/10.1126/scitranslmed.abd5487.10.1126/scitranslmed.abd5487Search in Google Scholar PubMed PubMed Central


Supplementary Material

The online version of this article offers supplementary material (https://doi.org/10.1515/cclm-2020-1758).


Received: 2020-11-25
Accepted: 2021-01-05
Published Online: 2021-01-15
Published in Print: 2021-05-26