Diagnostic accuracy of the ultrasensitive S-PLEX SARS-CoV-2 N electrochemiluminescence immunoassay

  • Giuseppe Lippi ORCID logo EMAIL logo , Brandon M. Henry , Martina Montagnana and Mario Plebani ORCID logo

To the Editor,

We read with interest the recent article of Ren et al. [1], who described the accuracy of an ultrasensitive electrochemiluminescence immunoassay for saliva-based Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) nucleocapsid protein (N) detection based on the S-PLEX platform (S-PLEX SARS-CoV-2 N Kit; Meso Scale Discovery, Rockville, MD, United States). This method has been specifically developed for detecting and quantifying the SARS-CoV-2 N antigen in a variety of human specimens, including serum, plasma, saliva and nasopharyngeal swabs (NPS). Briefly, either S-PLEX 96-Well SECTOR or QuickPlex plates coated with streptavidin for binding biotin-conjugated capture anti-SARS-CoV-2 N antibodies are challenged with human samples. After this step, “TURBO-BOOST”-labeled detection antibodies react with the N antigen bound to the solid phase and, after addiction of a specific reagent, an electrochemiluminescent signal is generated and read by the specific instrument. The signal produced is proportional to the concentration of N antigen present in the test sample. A preliminary evaluation of this assay revealed that the limit of detection is 0.16 pg/mL, with a diagnostic threshold set at 0.32 pg/mL and a total imprecision ranging between 7.0 and 7.7% [2]. The sample volume is only 25 μL, with total turnaround time between 4–5 h.

Since this novel technique displayed remarkable diagnostic performance in saliva samples in the hands of Ren and colleagues, exhibiting up to 100% specificity with 92% sensitivity [1], we provide here a critical literature review and pooled analysis of studies which addressed the accuracy of S-PLEX SARS-CoV-2 N Kit for diagnosing acute SARS-CoV-2 infections.

We carried out a digital search in the two scientific databases Medline (PubMed interface) and Scopus, using the following keywords: “S-PLEX” AND “COVID-19” OR “SARS-CoV-2”, without no language or date (i.e., up to February 17, 2022) restrictions. The initial screening of documents was conducted by G.L. and M.M., aimed at selecting studies were the diagnostic accuracy of S-PLEX SARS-CoV-2 N Kit was assessed against a reference molecular technique for diagnosing acute SARS-CoV-2 infections, and with sufficient extrapolable information for construction of a 2×2 table. A pooled analysis, based on the Mantel-Haenszel method and random effects model, was employed for estimating the diagnostic sensitivity, specificity and accuracy (reported as Summary Receiver Operating Characteristic Curve [SROC] and agreement) of this method. The inter-study heterogeneity was also assessed with χ2 test and I2 statistic. The statistical analysis was performed with Meta-DiSc 1.4 (Unit of Clinical Biostatistics team of the Ramón y Cajal Hospital, Madrid, Spain) [3]. The analysis was carried out in accordance with the Declaration of Helsinki and within the terms of local legislation. No Ethical Committee approval was necessary, as this is a systematic literature review.

Our electronic search identified six articles after duplicates elimination. One article was excluded as it did not present data on S-PLEX SARS-CoV-2 and another because containing insufficient information for constructing a 2×2 table, thus four studies (totaling 657 clinical samples) were included in our analysis (Table 1) [1, 2, 4, 5]. An additional search, using the key word “Meso Scale”, allowed to detect 12 more articles, but none of these presented data on S-PLEX SARS-CoV-2 N Kit.

Table 1:

Summary of studies which explored the diagnostic accuracy of S-PLEX SARS-CoV-2 N electrochemiluminescence immunoassay for diagnosing acute SARS-CoV-2 infections.

Study Country Study population Sample size Disease prevalence Sample Presumptive S-PLEX cut-off COVID-19 diagnosis
Pollock et al. [2] USA 85 adult and 141 pediatric patients; 65% females 226 60.1% (95%CI, 53.5–66.6%) NPS ≤0.32 pg/mL Multiple molecular assays in NPS
Ren et al. [1] Canada Non specified 105 48.0% (95%CI, 38.0–58.2%) Saliva ≤0.32 pg/mL In-house molecular assay in saliva
Wang et al. [4] USA Non specified 200 50.0% (95%CI, 42.9–57.1%) NPS ≤0.28 pg/mL Panther Fusion RT-PCR in NPS
Wang et al. [5] USA 57 (IQR, 39–72) years; 50% females 126 58.7% (95%CI, 49.6–67.4%) Plasma ≤0.32 pg/mL Multiple molecular assays in NPS
  1. 95%CI, 95% confidence interval; IQR, interquartile range; NPS, nasopharyngeal swab; RT-PCR, real-time polymerase chain reaction.

All four studies were from North America (three from the US and one from Canada), with a sample size comprised between 105 and 226 clinical specimens and cumulative disease prevalence of 53.4% (95%CI, 49.5–57.3%). The population from which the specimens were collected was only specified in two instances (Table 1). SARS-CoV-2 diagnosis was made by using multiple molecular assays in two cases, and a commercial and an in-house molecular technique in the remaining two studies. Overall, molecular detection of SARS-CoV-2 was carried out in NPS in three investigations and saliva in the remaining study. As concerns S-PLEX, SARS-CoV-2 antigen detection was conducted in NPS in two studies, and in saliva or plasma in the remaining two investigations. In three cases, a cut-off of ≤0.32 pg/mL was used, whilst the remaining investigation applied a different threshold.

The cumulative diagnostic accuracy of S-PLEX for diagnosing acute SARS-CoV-2 infections patients is summarized in Figure 1. The diagnostic sensitivity was 0.87 (95%CI, 0.83–0.90), the specificity 0.92 (95%CI, 0.89–0.95), the area under the curve (AUC) 0.955 (95%CI, 0.944–0.967), whilst the accuracy was 89.5% (95%CI, 86.9–91.7%). Similar results were obtained when our analysis was restricted to the three studies which evaluated SARS-CoV-2 in NPS or saliva, in that the sensitivity was 0.86 (95%CI, 0.81–0.89), the specificity 0.92 (95%CI, 0.88–0.95), the AUC 0.950 (95%CI, 0.934–0.966) and the accuracy 88.7% (85.7–91.3%), respectively.

Figure 1: 
Cumulative diagnostic accuracy of S-PLEX SARS-CoV-2 N electrochemiluminescence immunoassay for diagnosing acute SARS-CoV-2 infections. 95%CI, 95% confidence interval.
Figure 1:

Cumulative diagnostic accuracy of S-PLEX SARS-CoV-2 N electrochemiluminescence immunoassay for diagnosing acute SARS-CoV-2 infections. 95%CI, 95% confidence interval.

The results of our critical literature review and pooled analysis confirm that the ultrasensitive S-PLEX SARS-CoV-2 N electrochemiluminescence immunoassay displays excellent performance for detecting SARS-CoV-2 N antigen in a vast array of clinical specimens (NPS, saliva and even plasma), exhibiting a cumulative diagnostic sensitivity as high as 87%, which would make it one of the most sensitive SARS-CoV-2 antigen tests currently available in the market. As comparison, the cumulative diagnostic sensitivity of this method was found to be substantially better than that found with laboratory-based chemiluminescent techniques such as DiaSorin LIAISON SARS-CoV-2 Ag test (i.e., 31%) [6] and Roche Elecsys SARS-CoV-2 antigen electrochemiluminescence immunoassay (i.e., 68%) [7], slightly higher than that reported for Ortho VITROS SARS-CoV-2 antigen test (i.e., 82%) [8], and exactly identical to that of the other high-sensitivity immunoassay Fujirebio Lumipulse SARS-CoV-2 Ag test (i.e., 87%) [9].

In conclusion, the relatively low rate of false negative test results (i.e., 13%) in high-prevalence settings (i.e., disease prevalence around 50%) would make the ultrasensitive S-PLEX SARS-CoV-2 N electrochemiluminescence immunoassay a good surrogate of molecular testing for initial screening of patients with high clinical suspicion of SARS-CoV-2 infection. It is also important to mention here, that this method has also been validated for measuring SARS-CoV-2 N antigen in serum or plasma, thus paving the way to improve prognostication of COVID-19. In fact, the serum concentration of SARS-CoV-2 N protein is significantly associated with disease severity, as mirrored by higher risk of hospitalization and longer hospital stay [10], but also display high diagnostic sensitivity in specimens with high viral load (i.e., up to 100% in those with cycle threshold value <33) [11].


Corresponding author: Prof. Giuseppe Lippi, IFCC Task Force on COVID-19, Milan, Italy; and Section of Clinical Biochemistry, University of Verona, Piazzale L.A. Scuro, 10, 37134 Verona, Italy, Phone: +39 045 8122970, Fax: +39 045 8124308, E-mail:
Martina Montagnana and Mario Plebani share senior authorship of this work.
  1. Research funding: The authors received no funding for this work.

  2. Author contributions: All authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  3. Competing interests: Authors state no conflict of interest.

  4. Informed consent: Not applicable.

  5. Ethical approval: Not applicable.

References

1. Ren, A, Sohaei, D, Ulndreaj, A, Pons-Belda, OD, Fernandez-Uriarte, A, Zacharioudakis, I, et al.. Ultrasensitive assay for saliva-based SARS-CoV-2 antigen detection. Clin Chem Lab Med 2022;60:771–7. https://doi.org/10.1515/cclm-2021-1142.Search in Google Scholar PubMed

2. Pollock, NR, Savage, TJ, Wardell, H, Lee, RA, Mathew, A, Stengelin, M, et al.. Correlation of SARS-CoV-2 nucleocapsid antigen and RNA concentrations in nasopharyngeal samples from children and adults using an ultrasensitive and quantitative antigen assay. J Clin Microbiol 2021;59:e03077–20. https://doi.org/10.1128/JCM.03077-20.Search in Google Scholar PubMed PubMed Central

3. Zamora, J, Abraira, V, Muriel, A, Khan, KS, Coomarasamy, A. Meta-DiSc: a software for meta-analysis of test accuracy data. BMC Med Res Methodol 2006;6:31. https://doi.org/10.1186/1471-2288-6-31.Search in Google Scholar PubMed PubMed Central

4. Wang, H, Hogan, CA, Verghese, M, Solis, D, Sibai, M, Huang, C, et al.. SARS-CoV-2 nucleocapsid plasma antigen for diagnosis and monitoring of COVID-19. Clin Chem 2021;68:204–13. https://doi.org/10.1093/clinchem/hvab216.Search in Google Scholar PubMed PubMed Central

5. Wang, H, Hogan, CA, Verghese, M, Solis, D, Sibai, M, Huang, C, et al.. Ultra-sensitive Severe acute respiratory Syndrome Coronavirus 2 (SARS-CoV-2) antigen detection for the diagnosis of Coronavirus disease 2019 (COVID-19) in upper respiratory samples. Clin Infect Dis 2021;73:2326–8. https://doi.org/10.1093/cid/ciab063.Search in Google Scholar PubMed PubMed Central

6. Salvagno, GL, Gianfilippi, G, Fiorio, G, Pighi, L, De Nitto, S, Henry, BM, et al.. Clinical assessment of the DiaSorin LIAISON SARS-CoV-2 Ag chemiluminescence immunoassay. EJIFCC 2021;32:216–23.10.2139/ssrn.3834210Search in Google Scholar

7. Lippi, G, Henry, BM, Adeli, K. Diagnostic performance of the fully automated Roche Elecsys SARS-CoV-2 antigen electrochemiluminescence immunoassay: a pooled analysis. Clin Chem Lab Med 2022;60:655–61. https://doi.org/10.1515/cclm-2022-0053.Search in Google Scholar PubMed

8. Lippi, G, Nocini, R, Henry, BM. Critical literature review and pooled analysis of diagnostic accuracy of Ortho Vitros SARS-CoV-2 Antigen Test for diagnosing acute SARS-CoV-2 infections. J Med Biochem 2022;41:1–9.10.5937/jomb0-36107Search in Google Scholar

9. Caputo, V, Bax, C, Colantoni, L, Peconi, C, Termine, A, Fabrizio, C, et al.. Comparative analysis of antigen and molecular tests for the detection of Sars-CoV-2 and related variants: a study on 4266 samples. Int J Infect Dis 2021;108:187–9. https://doi.org/10.1016/j.ijid.2021.04.048.Search in Google Scholar PubMed PubMed Central

10. Zhang, Y, Ong, CM, Yun, C, Mo, W, Whitman, JD, Lynch, KL, et al.. Diagnostic value of nucleocapsid protein in blood for SARS-CoV-2 infection. Clin Chem 2021;68:240–8. https://doi.org/10.1093/clinchem/hvab148.Search in Google Scholar PubMed PubMed Central

11. Favresse, J, Bayart, JL, David, C, Didembourg, M, Gillot, C, Dogné, JM, et al.. Spike vs. nucleocapsid serum antigens for COVID-19 diagnosis and severity assessment. Clin Chem Lab Med 2022;60:e97–100. https://doi.org/10.1515/cclm-2021-1244.Search in Google Scholar PubMed

Received: 2022-02-19
Accepted: 2022-03-07
Published Online: 2022-03-14
Published in Print: 2022-05-25

© 2022 Walter de Gruyter GmbH, Berlin/Boston

Downloaded on 19.4.2024 from https://www.degruyter.com/document/doi/10.1515/cclm-2022-0155/html
Scroll to top button