Home Patient result monitoring of HbA1c shows small seasonal variations and steady decrease over more than 10 years
Article Open Access

Patient result monitoring of HbA1c shows small seasonal variations and steady decrease over more than 10 years

  • Niclas Rollborn , Kim Kultima and Anders Larsson EMAIL logo
Published/Copyright: June 12, 2024

Abstract

Objectives

Internal and external quality assurance materials often use highly processed matrixes. This can render the materials non-commutable. Monitoring laboratory methods with patient medians helps in identifying and correcting systematic errors that may affect diagnostic accuracy. The aim of the present study was to use HbA1c patient results for monitoring of method performance over time.

Methods

Test HbA1c results from 2010 to 2022 was analyzed (n=722,553) regarding changes over time and seasonal variation. The HbA1c testing was initially performed on a Cobas 501 instrument using immunological detection but in May 2017 the method was replaced by capillary electrophoresis on Capillarys 3 Tera.

Results

There was a steady decrease in HbA1c values. From 2011 to 2021 the decrease was for 0.10 percentile 6.6 %, lower quartile 7.9 %, median 10.2 %, mean values 9 %, upper quartile 11.2 %, and 0.90 percentile 9.3 %. No clear shift in HbA1c levels was observed due to the shift in methods. The median HbA1c values per month was approximately 44 mmol/mol (6.2 %, DCCT/NGSP). The only month with a median HbA1c that differed by more than 1 mmol/mol was July with a median value of 42 mmol/mol (6.0 %).

Conclusions

The patient data showed a similar decrease as in the National Diabetes Register which indicates that the method is stable over time without any sudden changes and that the seasonal variation is low. The continuous decrease in HbA1c values over time is most likely to a shift towards earlier detection of patient with diabetes and improved treatment.

Introduction

It is important that patient results are as accurate as possible. Thus, there is a need for quality control of the test results. The test results are usually monitored by internal controls usually applying the Westgard rules and external quality assurance (EQA) programs [1, 2]. The internal quality control materials need to have a long shelf life which means that they have been treated to ensure that they are stable over time, and they may also contain non-human additives [3]. This could lead to commutability problems [4, 5]. The problem for EQA programs is that there is a need for large sample volumes of stable materials that can be distributed to the laboratories [6]. The larger EQA programs have hundreds of participants, and each participant will receive 0.5–1 mL of sample which adds up to several hundred mL. In most cases it is not possible to obtain and ship such large volumes of fresh genuine human samples. The EQA providers thus often rely on outdated surplus human materials from routine testing. To increase the stability, the samples are often frozen or freeze dried and stabilizers are added. The problem for the EQA organizations and the laboratories is that the preparation may be commutable with some reagents/instruments but not all [2, 7]. The lack of commutability may lead to higher or lower assay results for some of the reagent-instrument combinations.

Even if internal quality controls and EQA samples will continue to be the basis for quality control of laboratory test results, there is a need for other ways to monitor our methods [8]. One such possibility is the use of continuous monitoring of the patient results [3, 9]. Current laboratory information systems are able to extract patient median/mean values for a specific analyte.

If there are sufficiently large number of samples analyzed, the center part of the test result distribution is usually stable over time providing that the analyte has a stable calibration. Median and mean values are usually more stable than e.g., the 2.5 and 97.5 percentiles. If the patient results show an upward or downward trend or shift, this should be further investigated regardless if the internal quality controls or EQA samples show the same pattern. Recently, the Swedish EQA organization Equalis started a new program for patient median values. The participants report regularly their patient median results for some of the high-volume analytes. This allows for interlaboratory comparisons.

Hemoglobin A1c (HbA1c) was used as a model for patient median value monitoring over time at our laboratory. There are several different methods available for measuring HbA1c such as boronate affinity chromatography, capillary zone electrophoresis (CZE), enzymatic assays, immunoassays, and ion exchange high-performance liquid chromatography (HPLC) [10, 11]. A previous study has shown good agreements between the clinically used HbA1c methods in Sweden [11]. There are also national quality targets for the analyte to ensure a low method variation [12]. HbA1c is a high-volume assay which assures a more accurate determination of patient median values. In 2010, the Swedish laboratories changed from a mono S calibration of HbA1c to the IFCC calibration in mmol/mol. Changes from one calibration to another is often accompanied with an increased risk of calibration problems. HbA1c results were collected from the start of reporting of IFCC calibrated test results by the laboratory and data was collected up to August 2022.

The aim of the study was to document any changes over time and to investigate possible seasonal variation on HbA1c results. The median value is the most frequently recommended variable to monitor. The aim of the study was also to compare the changes in median values with changes in mean and quartile values.

Materials and methods

Samples

Routine requested HbA1c samples at the Departments of Clinical Chemistry, Uppsala, were collected in K2-EDTA tubes (364664, BD Vacutainer Systems Plymouth, UK). The study period began when the laboratory changed to IFCC calibration in august 2010 and ended in august 2022 (n=722,553). The data was extracted without full patient identities and only the data of the sampling and the HbA1c value was included. 26 HbA1c results <20 (<4 %) or >130 mmol/mol (>14 %)were replaced with 20 (4 %) or 130 mmol/mol (14 %) (0.0036 % of the results). This study was approved by the Ethical Committee at Uppsala University (Dnr 01-367). The study was conducted in accordance with the Declaration of Helsinki (as revised in 2013).

Instruments

The HbA1c routine samples were initially analyzed immunologically on cobas c501 (Roche Diagnostics, Rotkreuz, Switzerland) using Tina-Quant reagent. From May 2017 the HbA1c method was changed to capillary electrophoresis on Capillarys 3 Tera (Cap3) from Sebia (Lisses, France). The instruments were participating in the external quality assurance program run by Equalis (Uppsala, Sweden).

Statistical calculations

Statistical analysis (linear ordinary regression incl. correlation coefficient) was performed and calculated by Excel 365 (Microsoft Corp, Seattle, WA, USA) and Statistica 10 (Tibco Software, Palo Alto, CA, USA).

Results

Changes in HbA1c results over time

The population consisted of 335,894 females (46.5 %) and 386,659 males (53.5 %). The median age was 65.6 years (interquartile range 53.1–74.5 years, total range 0.02–109 years). All HbA1c test results were sorted according to year from 2011 to 2021 (Figure 1 and Table 1). There was a slight steady decrease over time. From 2011 to 2021 the decrease was for mean values 9 %, median 10.2 %, lower quartile 7.9 %, upper quartile 11.2 %, 0.10 percentile 6.6 % and 0.90 percentile 9.3 %. The upper quartile and 0.90 percentile decreased over the study period by 7 mmol/mol (2.8 %) while the 0.10 percentile decreased by 2 mmol/mol (2.3 %) and the lower quartile decreased by 3 mmol/mol (2.4 %) (Table 1). No clear shift in HbA1c levels was observed due to the shift in methods. Mean, median, lower, and upper quartiles and 0.10 percentile and 0.90 percentile showed an even change over time. Max and min values showed a higher variation between years (Table 1).

Figure 1: 
HbA1c values over time. The data is presented as lower quartile, median, mean, and upper quartile for each year 2011–2021.
Figure 1:

HbA1c values over time. The data is presented as lower quartile, median, mean, and upper quartile for each year 2011–2021.

Table 1:

Number of HbA1c results reported per year and mean, median, min, max, lower, and upper quartiles and 0.10 and 0.90 percentiles for each year.

Valid n Mean Median Min Max Lower quartile Upper quartile Percentile 0.10 Percentile 0.90
2011 39,996 52 47.9 14 182 39.1 60.8 35.3 75
2012 43,956 51.6 47.3 11 159.5 38.6 60.9 34.8 75
2013 49,633 51.5 46.7 7.6 184 38 60.7 34.5 75.6
2014 53,105 50.4 45 4 189.9 37.1 59.8 33.5 75
2015 60,297 49.9 45 8 233.4 37 59 33.1 73.4
2016 66,009 48.9 44 13.2 186 36 57.6 32.2 72.9
2017 72,045 48 43 12.5 185 36 56 32.2 71
2018 73,889 47.7 43 12 195 36 55 33 69
2019 76,856 48 43 11 181 37 55 33 69
2020 70,850 47.2 43 3.7 191 36 54 33 68
2021 75,663 47.3 43 0 180 36 54 33 68
Difference 5 5 0 3 7 2 7

Seasonal variation of HbA1c results

July is the main vacation month in Sweden which coincided with the lowest number of HbA1c requests. During the other 11 months the median HbA1c values were around 44 mmol/mol (6.2 %) (range 44.0–44.8 mmol/mol), while the median value for July was 42 mmol/mol (6.0 %) (Table 1). The same pattern was also observed for mean, lower quartile and upper quartile values. The intermonth coefficient of variation (CV) was 0.67 % for mean, 1.45 % for median, 0.75 % for lower quartile and 1.08 % for upper quartile values. The median value for December 24–31 was the same as the median value for July, 42 mmol/mol (6.0 %).

Correlation between number of tests per month and mean HbA1c results

There was a correlation between the number of tests per month and the mean HbA1c value (Spearman R2=0.56) (Figure 2). The mean HbA1c decreased when the number of tests decreased.

Figure 2: 
Correlations between the number of tests per month and mean HbA1c value.
Figure 2:

Correlations between the number of tests per month and mean HbA1c value.

Discussion

This study demonstrates a practical approach to add patient data into the traditional laboratory QC program of internal and external control materials.

Monitoring laboratory methods using patient medians offers a more robust and clinically relevant approach, addressing the unique characteristics of patient populations and contributing to the overall quality and reliability of laboratory testing. Patient medians capture the central tendency of a patient population and is usually recommended as the primary measure for patient results as medians are more resistant to outliers than e.g., patient means (Figure 3).

Figure 3: 
Mean HbA1c values in the Swedish National Diabetes Register 2010–2022.
Figure 3:

Mean HbA1c values in the Swedish National Diabetes Register 2010–2022.

In this study the variation in patient HbA1c results was investigated. The Swedish National Diabetes Register (NDR) is a quality register containing patient information provided by healthcare providers nation-wide and is used as a tool for continuous quality assessment for diabetes care in Sweden [13]. NDR uses HbA1c as a quality marker for the diabetes care in Sweden [14, 15]. This registration allows the comparison between national trends for HbA1c with trends observed for the patient HbA1c values in this study. The parameters studied were mean, median, lower quartile, upper quartile values, 0.10 percentile, 0.90 percentile and min and max values. The inter-month coefficient of variation for the four studied parameters were similar, indicating that it apart from median HbA1c values it is also possible to use the other parameters for assay monitoring. The addition of lower and upper quartiles increases the HbA1c range that can be monitored to approximately 36–54 mmol/mol (5.4–7.1 %). Including 0.1, and 0.9 percentiles increased the monitoring range even further to 33–68 mmol/mol (5.2–8.4 %). No sudden shift in HbA1c levels was observed when changing from immunological detection using Cobas c501 to capillary electrophoretic detection using Capillarys 3. This is in agreement with the internal method validation performed before moving the method to the new platform and previous method comparison [10, 11].

Normally, patients sampled during holiday periods are sicker than the individuals sampled during regular months. In this study the opposite was found with a positive correlation between the number of tests per month and HbA1c values. Our interpretation of these findings is that the regular monitoring of diabetes patients is reduced during holiday periods e.g. July and the Christmas period. As HbA1c mainly is used for long-time monitoring of diabetes patients the testing can easily be postponed until after the vacations. In contrast, acute testing of patients with potential diabetes will remain constant thus contributing to a lower median HbA1c level during the vacation period in Sweden.

When comparing HbA1c results from the beginning of the study period with the last part of the study period a decline in HbA1c values was detected regardless of which of the monitoring parameters that was used. The decline is similar to the decline observed in the Swedish diabetes registry [16]. Controlling for sex and Type 1 diabetes duration, mean HbA1c decreased between 2011 and 2017 in all age cohorts in Sweden (p<0.001) [16]. Thus, the decline in our patient results is not due to a method deviation but reflects the Swedish general trend towards reduced HbA1c values nationally which is due to earlier detection and more effective treatment of patients with diabetes. The number of diabetes patients and especially type II diabetes patients are increasing worldwide [17]. Lowering HbA1c levels is often a primary objective in diabetes management because it correlates with reduced risk of long-term complications associated with diabetes. Advancements in diabetes care, including the development of new medications, glucose monitoring technologies, improved insulin formulations, and better education and support for patients, have contributed to better glycemic control for many individuals. Additionally, lifestyle modifications such as dietary changes, regular physical activity, and weight management can also lead to improvements in HbA1c levels. Regular monitoring, personalized treatment plans, and ongoing support from healthcare providers are crucial for optimizing diabetes care and achieving target HbA1c levels. Early detection and treatment will reduce chronic damage due to diabetes. The increasing number of test results per year and the tendency that the change over time was less pronounced for the 0.1 percentile and lower quartile supports the interpretation that the reduction was due to earlier detection of diabetes patient or more efficient treatment of diabetes patients leading to reduced HbA1c values especially in the upper HbA1c ranges.

Patient median monitoring can help in early detection of methodological issues that might affect test accuracy and precision. Timely identification allows for corrective actions to be taken before widespread issues occur. The patient median results are used for internal QC control. If results from the external quality assurance program indicates a problem, we compare the results with the patient median before we take any actions. We also use patient medians to verify that batch changes of reagents have not caused deviations in test results. The patient median can also be used to verify that there has not been any shift in calibration over time in clinical studies that are ongoing for several months/years.

Monitoring laboratory methods with patient medians helps in identifying and correcting systematic errors that may affect diagnostic accuracy. This ensures that test results are reliable and can be confidently used by healthcare professionals for patient diagnosis and treatment decisions. A limitation of the use of patient medians is that it is based on the assumption that the sampled patient groups are stable. If the number of assays per year is limited a shift in the use of a specific assay could influence the patient median values. For instance, creatinine is mainly used as a GFR marker in very large volumes. The patient median seems to be very stable. In contrast, the GFR marker cystatin C is sometime used in primary care exclusively for very old patients where creatinine is considered by the GPs as less reliable. This will create a bias when comparing patient medians with other hospitals. Also, changes over time e.g. reduced HbA1c median values during July as shown in this study could cause interpretation problems (Table 2).

Table 2:

Seasonal variation of HbA1c results for the time period 2010–2022. The results are sorted according to sampling month.

Valid n Mean Median Lower quartile Upper quartile
January 65,298 49.40 44.0 37.0 58.0
February 65,151 49.12 44.0 37.0 57.7
March 65,536 49.29 44.0 37.0 58.0
April 65,536 48.89 44.0 37.0 57.0
May 65,536 49.16 44.8 37.0 57.0
June 57,892 48.70 44.0 36.9 57.0
July 30,084 48.13 42.0 36.0 56.0
August 59,009 48.57 44.0 37.0 56.0
September 65,536 49.03 44.5 37.0 57.0
October 65,536 48.81 44.0 37.0 57.0
November 65,536 48.75 44.0 37.0 56.7
December 51,903 48.99 44.0 37.0 57.0

Conclusions

In conclusion, monitoring laboratory methods using patient medians, means, quartiles and 0.1 and 0.9 percentiles offers a more robust and clinically relevant approach to method monitoring. The data shows that the HbA1c has a small seasonal variation and that the HbA1c method has had a stable calibration between 2010 and 2022. The decline over time is similar to the decrease observed in the Swedish National Diabetes Register.


Corresponding author: Anders Larsson, Department of Medical Sciences, Section of Clinical Chemistry, Uppsala University, Uppsala, Sweden, Akademiska Sjukhuset, Entrance 61, 3rd floor SE-751 85, Uppsala, Sweden, E-mail:

Funding source: Uppsala University Hospital Research Fund

  1. Research ethics: The local Ethical Committee (01-367) approved the collection of samples. The study was conducted in accordance with the Declaration of Helsinki (as revised in 2013).

  2. Informed consent: Not applicable.

  3. Author contributions: The authors have accepted responsibility for the entire content of this manuscript and approved its submission.

  4. Competing interests: The authors state no conflict of interest.

  5. Research funding: The Uppsala University Hospital Research Fund supported this study.

  6. Data availability: The raw data can be obtained on request from the corresponding author.

References

1. Westgard, JO, Barry, PL, Hunt, MR, Groth, T. A multi-rule Shewhart chart for quality control in clinical chemistry. Clin Chem 1981;27:493–501. https://doi.org/10.1093/clinchem/27.3.493.Search in Google Scholar

2. Ricós, C, Fernandez-Calle, P, Perich, C, Westgard, JO. Internal quality control – past, present and future trends. Adv Lab Med 2022;3:243–62. https://doi.org/10.1515/almed-2022-0029.Search in Google Scholar PubMed PubMed Central

3. Wilson, A, Roberts, WL, Pavlov, I, Fontenot, J, Jackson, B. Patient result median monitoring for clinical laboratory quality control. Clin Chim Acta 2011;412:1441–6. https://doi.org/10.1016/j.cca.2011.04.024.Search in Google Scholar PubMed

4. Skitek, M, Martinello, F, Jerin, A. How to really understand and improve the system of internal quality control and external quality assessment in the accreditation process of the medical laboratory? EJIFCC 2022;33:23–7.Search in Google Scholar

5. Ricós, C, Juvany, R, Alvarez, V, Jiménez, CV, Perich, C, Minchinela, J, et al.. Commutability between stabilized materials and fresh human serum to improve laboratory performance. Clin Chim Acta 1997;263:225–38. https://doi.org/10.1016/s0009-8981(97)00062-4.Search in Google Scholar PubMed

6. Badrick, T, Punyalack, W, Graham, P. Commutability and traceability in EQA programs. Clin Biochem 2018;56:102–4. https://doi.org/10.1016/j.clinbiochem.2018.04.018.Search in Google Scholar PubMed

7. Rej, R. Accurate enzyme activity measurements. Two decades of development in the commutability of enzyme quality control materials. Arch Pathol Lab Med 1993;117:352–64.Search in Google Scholar

8. Jørgensen, LM, Hansen, SI, Petersen, PH, Sölétormos, G. Median of patient results as a tool for assessment of analytical stability. Clin Chim Acta 2015;446:186–91. https://doi.org/10.1016/j.cca.2015.04.024.Search in Google Scholar PubMed

9. Lu, Y, Yang, F, Wen, D, Shi, K, Gu, Z, Lu, Q, et al.. Assessment of patient based real-time quality control on comparative assays for common clinical analytes. J Clin Lab Anal 2022;36:e24651. https://doi.org/10.1002/jcla.24651.Search in Google Scholar PubMed PubMed Central

10. Rollborn, N, Åkerfeldt, T, Nordin, G, Xu, XY, Mandic-Havelka, A, Hansson, LO, et al.. Analysis of HbA1c on an automated multicapillary zone electrophoresis system. Scand J Clin Lab Invest 2017;77:15–18. https://doi.org/10.1080/00365513.2016.1238507.Search in Google Scholar PubMed

11. Vallée, M, Mandic-Havelka, A, Anders, L, Elmgren, A, Benny, L, Sisowath, C, et al.. Good agreement between HbA1c analyzed using capillary electrophoresis, HPLC, immunological and enzymatic methods. J Diabetes Metabol Complicat 2019;1:1–7. https://doi.org/10.31487/j.jdmc.2019.01.03.Search in Google Scholar

12. Nordin, GA-O. Accuracy of HbA1c as monitored by external quality assessment and compared with patient mean values. J Diabetes Science Technol 2018;12:771–9.10.1177/1932296818785622Search in Google Scholar PubMed PubMed Central

13. Gudbjörnsdottir, S, Cederholm, J, Nilsson, PM, Eliasson, B. The national diabetes register in Sweden: an implementation of the St. Vincent declaration for quality improvement in diabetes care. Diabetes Care 2003;26:1270–6. https://doi.org/10.2337/diacare.26.4.1270.Search in Google Scholar PubMed

14. Hallgren Elfgren, IM, Grodzinsky, E, Törnvall, E. Swedish Diabetes Register, a tool for quality development in primary health care. Prim Health Care Res Dev 2013;14:250–7. https://doi.org/10.1017/s1463423612000515.Search in Google Scholar

15. Hallgren Elfgren, IM, Grodzinsky, E, Törnvall, E. The Swedish National Diabetes Register in clinical practice and evaluation in primary health care. Prim Health Care Res Dev 2016;17:549–58. https://doi.org/10.1017/s1463423616000098.Search in Google Scholar PubMed

16. Albanese-O’Neill, A, Grimsmann, JM, Svensson, AM, Miller, KM, Raile, K, Akesson, K, et al.. Changes in HbA1c between 2011 and 2017 in Germany/Austria, Sweden, and the United States: a lifespan perspective. Diabetes Technol Therapeut 2022;24:32–41. https://doi.org/10.1089/dia.2021.0225.Search in Google Scholar PubMed

17. GBD 2021 Diabetes Collaborators, Stafford, LK, McLaughlin, SA, Boyko, EJ, Vollset, SE, Smith, AE, et al.. Global, regional, and national burden of diabetes from 1990 to 2021, with projections of prevalence to 2050: a systematic analysis for the Global Burden of Disease Study 2021. Lancet 2023;402:203–34. https://doi.org/10.1016/s0140-6736(23)01301-6.Search in Google Scholar

Received: 2024-02-12
Accepted: 2024-05-14
Published Online: 2024-06-12
Published in Print: 2024-11-26

© 2024 the author(s), published by De Gruyter, Berlin/Boston

This work is licensed under the Creative Commons Attribution 4.0 International License.

Articles in the same Issue

  1. Frontmatter
  2. Editorial
  3. External quality assurance (EQA): navigating between quality and sustainability
  4. Reviews
  5. Molecular allergology: a clinical laboratory tool for precision diagnosis, stratification and follow-up of allergic patients
  6. Nitrous oxide abuse direct measurement for diagnosis and follow-up: update on kinetics and impact on metabolic pathways
  7. Opinion Papers
  8. A vision to the future: value-based laboratory medicine
  9. Point-of-care testing, near-patient testing and patient self-testing: warning points
  10. Navigating the path of reproducibility in microRNA-based biomarker research with ring trials
  11. Point/Counterpoint
  12. Six Sigma – is it time to re-evaluate its value in laboratory medicine?
  13. The value of Sigma-metrics in laboratory medicine
  14. Genetics and Molecular Diagnostics
  15. Analytical validation of the amplification refractory mutation system polymerase chain reaction-capillary electrophoresis assay to diagnose spinal muscular atrophy
  16. Can we identify patients carrying targeted deleterious DPYD variants with plasma uracil and dihydrouracil? A GPCO-RNPGx retrospective analysis
  17. General Clinical Chemistry and Laboratory Medicine
  18. Comparison of ChatGPT, Gemini, and Le Chat with physician interpretations of medical laboratory questions from an online health forum
  19. External quality assessment performance in ten countries: an IFCC global laboratory quality project
  20. Multivariate anomaly detection models enhance identification of errors in routine clinical chemistry testing
  21. Enhanced patient-based real-time quality control using the graph-based anomaly detection
  22. Performance evaluation and user experience of BT-50 transportation unit with automated and scheduled quality control measurements
  23. Stability of steroid hormones in dried blood spots (DBS)
  24. Quantification of C1 inhibitor activity using a chromogenic automated assay: analytical and clinical performances
  25. Reference Values and Biological Variations
  26. Time-dependent characteristics of analytical measurands
  27. Cancer Diagnostics
  28. Expert-level detection of M-proteins in serum protein electrophoresis using machine learning
  29. An automated workflow based on data independent acquisition for practical and high-throughput personalized assay development and minimal residual disease monitoring in multiple myeloma patients
  30. Cardiovascular Diseases
  31. Analytical validation of the Mindray CL1200i analyzer high sensitivity cardiac troponin I assay: MERITnI study
  32. Diabetes
  33. Limitations of glycated albumin standardization when applied to the assessment of diabetes patients
  34. Patient result monitoring of HbA1c shows small seasonal variations and steady decrease over more than 10 years
  35. Letters to the Editor
  36. Inaccurate definition of Bence Jones proteinuria in the EFLM Urinalysis Guideline 2023
  37. Use of the term “Bence-Jones proteinuria” in the EFLM European Urinalysis Guideline 2023
  38. Is uracil enough for effective pre-emptive DPD testing?
  39. Reply to: “Is uracil enough for effective pre-emptive DPD testing?”
  40. Accurate predictory role of monocyte distribution width on short-term outcome in sepsis patients
  41. Reply to: “Accurate predictory role of monocyte distribution width on short-term outcome in sepsis patients”
  42. Spurious parathyroid hormone (PTH) elevation caused by macro-PTH
  43. Setting analytical performance specifications for copeptin-based testing
  44. Serum vitamin B12 levels during chemotherapy against diffuse large B-cell lymphoma: a case report and review of the literature
  45. Evolution of acquired haemoglobin H disease monitored by capillary electrophoresis: a case of a myelofibrotic patient with a novel ATRX mutation
Downloaded on 24.6.2025 from https://www.degruyterbrill.com/document/doi/10.1515/cclm-2024-0205/html
Scroll to top button